Shenzhen Pray-med Technology Co.,Ltd

Shenzhen Pray-med Technology Co.,Ltd,Heart with sincerity, good service every customer is our highest goal

Manufacturer from China
Verified Supplier
8 Years
Home / Products / Medical Oxygen Sensor /

Ventilator Medical O2 Sensor , Quick React Original New O2 Sensor Medical

Contact Now
Shenzhen Pray-med Technology Co.,Ltd
Visit Website
City:shenzhen
Province/State:guangdong
Country/Region:china
Contact Person:Mrssarah
Contact Now

Ventilator Medical O2 Sensor , Quick React Original New O2 Sensor Medical

Ask Latest Price
Brand Name :drager
Model Number :6850645
Certification :CE,ISO
Place of Origin :Germany
MOQ :1pcs
Price :as discussed
Payment Terms :T/T, Western Union,
Supply Ability :300pcs / week
Delivery Time :5-7 work days
Packaging Details :1pcs / bag
COLOR :silver
brand :drager
material :Mixture
warranty :12 months
name :o2 sensor medical
approval :CE,ISO
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Original and new drager medical Oxygen sensor 6850645 for Ventilator 6850645 O2 sensor O2 cell

Sensor Specifications & Compatibility
Catalog Number
Output Signal in Air 24A - 32A (Dual Cathode) no temperature compensation
Response Time T 90% < 12 seconds
Nominal Sensor Lifetime 500 000 % volume Oxygen hours
Electrical Interface Gold plated slip rings
Compatibility: Draeger 6850645 (Dr¤ger)
Babylog, Fabian, Cicero, PM8030
Evita, Oxydig, Anemone, Oxycom
Incubator 8000
M08 ST13
MAX11
Invivo Magnitude
10-103-07
PT14A
Teledyne R13 C-61139
Hudson 5557 (V14)
VTI 101372, 100352
* Note: Order E61720 with reusable cap and "o" rings
Medical Oxygen Sensors (Fuel Cells) general specifications:
Measurement Range: 0-100 % oxygen
Accuracy and Repeatability: < 1 % vol. O2when calibrated at 100 % O2
Zero Ofset: < 0.5 % vol. O2in 100% N2, applied 5 minutes
Linearity Error: < 3 % relative
Cross Interference: < 0.5 % vol. O2response to:
10 % CO2balance N2
80% N2O balance N2
7.5% Halothane balance N2
7.5 % Isofurane balance N2
7.5 % Enfurane balance N2
9% Sevofurane balance N2
20% Desfurane balance N2
Infuence of Humidity: - 0.03 % rel. per % RH at 25C
Infuence of Pressure: proportional to change in oxygen partial
pressure
Infuence of Mechanical Shock: < 1% relative after a fall from 1m
Operating Temperature: 0 to 50C
Temperature Compensation: built-in NTC compensation (see below)
Efect of Temperature between +5 C and +25 C: 3 % relative error
Compensation (steady state): between 0 C and +50 C: 8 % relative error
Operating Humidity: 0-99 % RH non-condensing
Long Term Output Drift: < 1 % vol oxygen per month
typically < - 15 % relative over lifetime
Storage Temperature: -20 to +50 C
Recommended Storage: +5 to +15 C
Recommended Load: 10 kOhms
Warm-Up Time: < 30 minutes, after replacement of sensor
Weight: approximately 28 grams
approximately 43 grams E61912 series
Designed and manufactured according to EN ISO 9001 : 2000 and EN13485 : 2003
specifications are applicable at standard conditions: 1013 hPa, 25C dry ambient air
Oxygen sensor for Draeger OEM 6850645
Envitec 6850645

Ventilator Medical O2 Sensor , Quick React Original New O2 Sensor Medical

Inquiry Cart 0